Mechanical Vibration - A quick and EfficienMethod of Removing Damaged Prosthesis from its Cemented Position
محتوى المقالة الرئيسي
الملخص
Implant replacement or revision surgery in the
human body ranks among the most demanding surgeries
that orthopedic surgeons encounter in modern medicine.
The removal of the hardened bone cement which secures
the implant within the bone, is the cause of a lengthy
operation. Often the use of sharp surgical instruments,
ultrasonic devices etc. are employed, however all these
methods are time-consuming and present various
complications.
This paper describes the use of mechanical vibrations at a
frequency of 40 Hz, on samples of implants anchored in
animal (pig) bone cavities with bone cement. Using a rotary
hammer drill equipped with a specially designed adaptor, to
engage the tip of the implant, good results were obtained.
The implant with its cement mantle was removed in a short
time not exceeding 22 seconds.
The procedure appears to be safe and is accomplished in
a short time, commensurate with what is required in
revision surgery.
تفاصيل المقالة
هذا العمل مرخص بموجب Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
المراجع
M. Vert et al ., ‘Terminology for biorelated polymers
and applications (IUPAC Recommendations 2012)’, Pure Appl.
Chem., vol. 84, no. 2, p. 1, 2012.
R. M. Guedes, M. Gomes, and J. A. Simões, ‘DMTA analysis for
long-term mechanical behaviour prediction of PMMA-based bone
cements’, J. Biomater. Sci. Polym. Ed., vol. 17, no. 10, pp. 1173–
, 2006.
P. N. Smith and K. S. Eyres, ‘Safe removal of massive intrapelvic
cement using ultrasonic instruments’, J. Arthroplasty, vol. 14, no.
, pp. 235–238, 1999.
R. Gardiner, W. J. Hozack, C. Nelson, and E. M. Keating,
‘Revision total hip arthroplasty using ultrasonically driven tools:
a clinical evaluation’, J. Arthroplasty, vol. 8, no. 5, pp. 517–521,
NICE, ‘ The OSCAR 3 ultrasonic arthroplasty revision
instrument for removing bone cement during prosthetic joint
revision,’ November 19, pp. 1–22, 2014
https://www.nice.org.uk/advice/mib13 last visited 13 Jul 2019
A. T. Brooks, C. Nelson, C. L. Stewart, R. A. Skinner, and
M . L. Siems, ‘ Effect of an ultrasonic device on
temperatures generated in bone and on bone-cement structure.’,
J.Arthroplasty, vol. 8, no. 4. pp. 413–418, 1993 DOI: https://doi.org/10.1016/S0883-5403(06)80041-0
P. Postawa and A Szarek ‘Analysis of changes in bone cement
damping factor and its effect on bone load’, vol 23, pp . 35–38,